Model Number
Temperature Sensor Measurement Range
Operating Temperature Range
Storage Temperature Range
Ambient humidity range during transport and
use
Rated Temperature Resolution
Rated Temperature Accuracy
Probe Compatibility
Sensor Reaction Time
Clock accuracy
Recording Capacity
Statistics memory
Memory type
Sampling Interval
Start options
Alarm functions
Vibration
Shock
EMC compliance
Environmental
Case Material
Power source
UTREL-16 User Guide
UTREL-16
-90 °C to +40 °C (-130 °F to +104 °F)
-30 °C to +70 °C (-22 °F to +158 °F)
0 °C to +40 °C (32 °F to +104 °F)
0 to 95 %RH
0.1 °C (0.2 °F) for -85 °C to 0 °C (-121 °F to +32 °F)
0.2 °C (0.4 °F) for 0 °C to +18 °C (+32 °F to +64 °F) and -90 °C to -85 °C (-130 °F to -121
°F)
0.3 °C (0.5 °F) for +18 °C to +30 °C (+64 °F to +86 °F)
0.5 °C (0.9 °F) for +30 °C to +40 °C (+86 °F to +104 °F)
Please see resolution chart in
Accuracy and Resolution
Better than ±0.5 °C (±0.9 °F) for -20 °C to +10 °C (-4 °F to 50 °F)
Better than ±0.7 °C (±1.3 °F) for -50 °C to -20 °C (-58 °F to -4 °F) and +10 °C to +30 °C
(+50 °F to +86 °F)
Better than ±1.0 °C (±1.8 °F) for -85 °C to -50 °C (-121 °F +-58 °F) and +30 °C to +40 °C
(+86 °F to +104 °F)
Better than ±1.8 °C (±3.2 °F) for -90 °C to -85 °C (-130 °F to -121 °F)
Please see accuracy chart in
Accuracy and Resolution
Any external probe from the ST100 series
According to Probe
Quartz crystal-locked real time clock, rated accuracy ±25ppm @ 25 °C (equiv to 2.5
seconds/day)
Rated temperature coefficient is -0.034±0.006ppm/°C (i.e. typically +/-
0.00294seconds/day/°C)
16,129 real time temperature values, giving
112 days @ 10 min logging, or
l
168 days @ 15 min logging
l
For displaying statistics on the LCD
trip min/max values
l
Total time above/below alarm thresholds
l
Non volatile
Configurable from 30 seconds to 18 hours
Push button start with optional configurable start delay from 1 minute to 72 hours
l
Date/time start
l
OK tick and Alarm cross on display, linked to alarms
l
Alarm arrows linked to thresholds
l
three configurable upper alarms
l
three configurable lower alarms
l
Withstands vibration specification as detailed in EN12830:2018
Withstands shock specification as detailed in EN12830:2018
l
Withstands 5 drops from 1m onto smooth concrete floor without loss of function or
l
calibration
EC EMC directives (CISPR 11: 2009 + A1: 2010, IEC 61000-4-3:2010, IEC 61000-4-
l
4:2012, IEC 61000-4-6:2008)
RTCA DO-160G:2010, section 21
l
Includes electrostatic discharge as prescribed in IEC 61000-4-2:2008
l
Complies with FCC Part 15 Subparts A and B
l
IEC 60529: IP61 with USB cap fitted and hung vertically
Polycarbonate
CR2032 3V LiMnO
coin cell battery, non-rechargeable, non-replaceable
2
44