3 Configuration
3.5
Set automatic power-off
3.6
Adr.
12
min
Use the
or
max
2
5
rH
mV
mV
pH
H
max
2
C
min
5
ATC
ok
CAL
AL Logg
pH
-10.0°C to +10.0°C
or
-18.0°F to +18.0°F
Off
Start configuration.
Select the parameter "P.oFF".
min
Use the
or
max
5
2
rH
mV
mV
pH
H
max
2
mbar
min
5
ATC
ok
CAL
AL Logg
pH
1 — 120:
Off
This function is not supported by the present instrument.
rH
mV
mV
pH
H
max
2
min
5
ATC
ok
CAL
AL Logg
pH
01, 11, 21, — 91:
key to set the temperature shift (offset).
The zero point for temperature measurement can be
shifted by this amount, to compensate for probe or
measurement deviations.
Measured temperature minus offset = displayed
temperature.
The zero point shift is switched off (0.0°).
key to set the power-off delay in minutes.
Automatic power-off delay in minutes.
If no key is pressed on the hand-held meter, it will switch
off after this time has elapsed.
Continuous operation (appropriate when fed from the
electrical supply).
No function