JUMO 202710 Operating Instructions Manual page 31

Hand-held meter
Hide thumbs Also See for 202710:
Table of Contents

Advertisement

Weight
Temperature
compensation
pH calibration
Recommended
redox
combination
electrodes
EMC
approx. 145 g
In the operating modes "pH", "mV
compensation (ATC) is possible with the aid of a Pt100 temperature sensor.
If there is no temperature sensor, the temperature of the medium can be
entered manually.
- 2-point or 3-point calibration
- Standard or DIN series buffers will automatically be detected and their
temperature dependency is automatically compensated.
- Manual buffers can be used
- Permissible electrode data: asymmetry: ±60 mV
- Sensor weighting (10 to 100%) according to the calibration results.
Reference system: Ag/AgCl, electrolyte: KCl, 3 mol/liter
JUMO ecoLine Rd, 2 m fixed connecting cable and BNC connector
Type: 201010/54-22-65-2000-76-120/000
Sales No. 20/00424950
(use only this type for conversion to the hydrogen system "mV
measurements!)
as per EN 61 326
interference emission: Class B
interference immunity: to industrial requirements
9 Technical data
" or "rH", automatic temperature
H
slope:
-62 to -45 mV/pH
" and rH
H
31