Mounting and sensor wiring
March 2020
2. Gently lift the transmitter straight up, and rotate the transmitter to the desired
3. Gently lower the transmitter onto the base, confirming that the transmitter is in a
4. Replace the clamp in its original position and tighten the cap screw. Torque to 28 in
24
position.
You can rotate the transmitter to any of the eight positions, but a stop exists that
will not allow a full 360° rotation.
Figure 3-15: Rotation of the transmitter head
locked position.
lbf (3.16 N m) to 30 in lbf (3.39 N m).
Figure 3-16: Re-attachment of the sensor clamp
Installation Manual
MMI-20029768
Micro Motion 5700 Ethernet Transmitters