Chapter 5 To Trigger the Oscilloscope
Setup Type:
It is used to select the different trigger types of setup/hold trigger. Press SetupType
to select the desired setup type.
Trigger Mode:
Press Sweep to open the trigger mode list and select auto, normal or single. For the
details, please refer to "Trigger Mode". The corresponding status light of the
current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameter (noise rejection) under this trigger type.
Trigger Level:
Press DataSource and use TRIGGER
data source channel. Press ClkSource and use TRIGGER
trigger level of the clock source channel. For the details, please refer to the
introduction of "Trigger Level".
DS1000Z-E User Guide
Find Quality Products Online at:
Setup: the oscilloscope triggers when the setup time is less than the specified
setup time. After selecting this type, press Setup to set the setup time and the
range is from 8 ns to 1 s.
Hold: the oscilloscope triggers when the hold time is less than the specified
hold time. After selecting this type, press Hold to set the hold time and the
range is from 8 ns to 1 s.
StpHold: the oscilloscope triggers when the setup time and hold time are less
than the specified times. After selecting this type, press Setup and Hold to set
the setup time and hold time respectively and the range is from 8 ns to 1 s.