Chapter 15 Troubleshooting

Chapter 15 Troubleshooting
The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still,
information (Utility  System  System Info).
1. The screen is still dark (no display) after you powered on the
2. The signal is sampled but no waveform of the signal is displayed.
3. The tested voltage amplitude is greater or smaller than the actual
4. Waveforms are found to be displayed but not stable.
5. No display is found after pressing RUN/STOP.
DS1000Z-E User Guide
Find Quality Products Online at:
instrument.
(1) Check whether the power switch is really on.
(2) Check whether the power is correctly connected.
(3) Check whether the fuse is burned out. If the fuse needs to be changed,
please use the specified fuse.
(4) Restart the instrument after finishing the above inspections.
(5) If it still persists,
(1) Check whether the probe is correctly connected to the oscilloscope and
item under test.
(2) Check whether signals are generated from the item under test (you can
connect the probe compensation signal to the problematic channel to locate
the problem).
(3) Resample the signal.
value (this problem usually occurs when probe is used).
Check whether the probe ratio of the channel complies with the attenuation ratio
of the probe.
(1) Check the trigger source: press MENU (in the trigger control area
(TRIGGER) on the front panel)  Source to check whether the trigger
source selected complies with the signal channel actually used.
(2) Check the trigger type: general signals should use "Edge" trigger and video
signal should use "Video" trigger. Only when the proper trigger type is used,
can the waveform be displayed stably.
(3) Check the trigger level: adjust the trigger level to the middle of the signal.
(4) Change the trigger holdoff setting.
Check whether the trigger mode in the trigger control area (TRIGGER) on the
front panel is "Normal" or "Single" and whether the trigger level exceeds the
waveform range. If yes, set the trigger level to the middle or set MODE to
"Auto".
Note:
Pressing AUTO could automatically finish the above setting.