6. The display of waveform is ladder-like.
(1) The horizontal time base might be too low. Increase the horizontal time
(2) If the display type is "Vectors", the lines between the sample points may
7. I fail to connect PC or PictBridge printer through USB.
(1) Press Utility  IO Setting  USB Device to check whether the current
(2) Check whether the USB cable is correctly connected to the oscilloscope and
(3) Check whether the USB cable is in good condition. If needed, restart the
8. The USB storage device cannot be recognized.
(1) Check whether the USB storage device can work normally.
(2) Make sure that the USB storage device being used is flash memory type and
(3) Make sure whether the capacity of the USB storage device is too large. It is
(4) Restart the instrument and then insert the USB storage device to check it.
(5) If the USB storage device still cannot be used normally, please contact
15-2
Find Quality Products Online at:
base to increase the horizontal resolution and improve the display.
cause ladder-like display. Press Display  Type to set the display type to
"Dots" to solve the problem.
setting matches the device currently connected.
PC.
oscilloscope.
FAT32 format. This oscilloscope does not support USB3.0 and hardware
type USB storage device.
recommended that the capacity of the USB storage device being used with
this oscilloscope should not exceed 8 GBytes.