Quick Calibration - Cardinal 212 Series Installation, Technical And Operation Manual

Table of Contents

Advertisement

Quick Calibration

Ag bar
cells=
int=
(Interval Setting)
8200-M639-O1 Rev D 09/14
1. Press the ON/OFF key to turn the indicator ON.
2. Press the F key (display will show
3. The display will change to the QUICAL prompt.
4. With QUICAL displayed, press the ENTER key. The display will change
to show the current setting NO.
5. Press the 1/YES key and then press the ENTER key to continue to the
AG BAR prompt.
(Ag Weigh Bar Selection)
Press the ENTER key. The current setting will be displayed.
1.
If the setting displayed is acceptable, press the ENTER key again to
2.
save it and advance to the CELLS= prompt.
Otherwise, using the numeric keys enter a new load cell number
3.
(allowable numbers are 1 to 22) from the Quick Calibration Weigh Bar
Table on page 43 and then press the ENTER key to save it.
The display will change to the CELLS= prompt.
4.
(Number of Load Cells)
Press the ENTER key to show the current setting.
1.
If the setting displayed is acceptable, press the ENTER key again to
2.
save it and advance to the iNT= prompt.
Otherwise, use the numeric keys enter a new setting (allowable
3.
settings for are 1 to 10) and then press the ENTER key to save it.
The display will change to show the Int= prompt.
4.
1. Press the ENTER key to show the current setting.
2. If the setting displayed is acceptable, press the ENTER key again to
save it and advance to the CAP= prompt.
3. Otherwise, use the numeric keys to enter a new setting (allowable
settings are: 1 through 99) and then press the ENTER key to save it.
4. The display will change to the CELLS= prompt.
212/212X Installation & Technical
F—
) and then press the 0/NO key.
Page 41