Storing Indicator Defaults

Restoring Indicator Defaults

8200-M639-O1 Rev D 09/14
1. Enter the Setup menu either by pressing the key sequence F, 0, .
2. With SETUP displayed, press the MEM key.
3. The indicator will briefly display SAVING then return to the Setup menu.
4. At this time, all setup information not related to calibration will be
stored as the default settings.
1. Enter the Setup menu either by pressing the key sequence F, 0, .
2. With SETUP displayed, press the ZERO/CLEAR key.
3. The indicator will display the prompt DFLTS? (Restore Defaults?).
4. Press the ENTER key to show the current setting.
5. If the setting displayed is acceptable, press the ENTER key again to
save it. Otherwise, using the numeric keys, 0/NO or 1/YES, enter the
new setting and then press the ENTER key to save it.
6. If YES was selected, the indicator will display LODING then return to the
Setup Menu.
a. If the defaults have been previously saved as described in the
previous section, these saved defaults will be restored
(RECALIBRATION IS NOT REQUIRED).
b. If no defaults have been previously stored, the indicator will restore
the new indicator defaults (RECALIBRATION IS REQUIRED).