2.4 INDICATOR OPERATIONAL SETUP REVIEW

Accessing and Navigating the Operational Setup Review

Operational Setup Review Parameters
8200-M639-O1 Rev D 09/14
This section describes another Setup Review that will only prompt for
operational parameters (e.g., DFC parameters, digital filtering).
1. With the indicator ON, press the F key followed by the PRESET key.
2. At each prompt, press the ENTER key to display the current value
associated with that prompt.
3. To modify any current value, enter the desired value and then press
the ENTER key.
4. Otherwise, press the ENTER key to keep the current value and
advance to the next prompt.
5. To step past a group of parameters, press the F key.
6. To step backwards to a previous prompt, press the  key.
7. To exit this menu, press the ENTER key after the last prompt or press
the  key at the first prompt.
Digital Output Control Enable/Disable and Settings
D oUt=
The following prompts are only displayed when D-Out= 4 or 14
DFC mode of operation
FMODE=
Fill Auto Accumulate
Fil-aa
Number of Digital Fill Cycles before purging feeder
purge=
Purge Weight – The amount of material contained in
p-wgt=
feeder before purge.
The next prompt is only displayed when D-Out= 1 or 11
Print On Balance
P-BAL=
The next prompt is only displayed when DOUT= 4 or 14 and FMODE= 0
Single or Dual speed DFC operation
SPEED=
212/212X Installation & Technical
Page 97