1-1. INTRODUCTION.
1-2. This
operating
and
service manual
contains
the information
required to install, operate, test,
adjust
and
service
the
Hewlett-Packard
Model
4262A Digital LCR Meter. Figure 1-1 shows the in-
strument
and supplied
accessories.
This section
covers
specifications,
instrument
identification,
description,
options,
accessories,
and
other basic
information.
1-3. Listed on the title page of this manual is a
~ tnicrofiche part number. This number can
be used
to order 4 x 6 inch microfilm transparencies of the
manual. Each microfiche contains up to 60 photo-
duplicates
of the manual
pages.
The
microfiche
package
also includes
the latest manual
changes
supplement
as well as all pertinent service notes.
To order an additional manual, use the part num-
ber listed on the title page of this manual.
1-4. DESCRIPTION.
1-5. The HP Model 4262A LCR Meter is a general
P/N 8120-1378
purpose, fully automatic test instrument designed
to measure
the parameters
of an impedance ele-
ment
with high accuracy and speed. The 4262A
measures capacitance, inductance, resistance (equi-
valent series resistance) and dissipation factor or
quality factor over a wide range at test frequencies
of 120Hz, 1kHz and 10kHz employing a five-term-
inal connection
configuration
between
the com-
ponent and the instrument.
The measuring circuit
for the device to be measured
is capable of both
parallel and series equivalent circuit measurements
and the measured values are displayed by the two
three-full digits LED displays on the front panel.
A convenient diagnostic function, also featured in
the 4262A,
is actuated by a SELF TEST switch.
This confirms functional operation of the instrument.
1-6. The measuring range for capacitance is from
0.01pF to 19.99mF,
inductance
from 0.014H
to
1999H,
and
resistance
from
ImQ
to 19.99M2,
which
are measured with a basic accuracy of 0.2
to 0.3% depending on test signal level, frequency,
and measuring
equivalent circuit,
and at typical
measuring speeds
of 220
to 260
milliseconds
at
5335
Ki
ain»
pn
cat
2;
Figure 1-1. Model 4262A and Accessories.