Trace Graticule Display; Trace X-Axis Scale Offset; Set The Display Line - HP ESA-E Series Programmer's Manual

Esa spectrum analyzers
Table of Contents

Advertisement

Language Reference
DISPlay Subsystem

Trace Graticule Display

:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Turns the graticule on or off.
Factory Preset
and *RST:
On
Front Panel
Access:
Display, Preferences, Graticule On Off

Trace X-Axis Scale Offset

:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet <freq>
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet?
Specifies the frequency offset for all frequency readouts such as center
frequency, except that it does not affect marker count.
Factory Preset
and *RST:
0 Hz
Range:
–3 GHz to 500 GHz
Default Unit:
Hz
Front Panel
Access:
FREQUENCY/Channel, Freq Offset

Set the Display Line

:DISPlay:WINDow:TRACe:Y:DLINe <ampl>
:DISPlay:WINDow:TRACe:Y:DLINe?
Defines the level of the display line, in the active amplitude units if no
units are specified.
Factory Preset
and *RST:
2.5 divisions below the reference level
Range:
10 display divisions below the reference level to the
reference level
Default Unit:
Current active units
Front Panel
Access:
Display, Display Line On Off
5-40
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading