8. Installation Verification

Power the instrument on and verify that the instrument will not only boot up, but that the signal
analyzer application software will load properly. This would include verifying that there are no error
messages displayed in the Status Bar of the application window.
Adjustments and Performance Verification Tests
Adjustments and performance verification testing requires the use of the calibration software. The
latest software information and downloads are available at:
http://www.agilent.com/find/calibrationsoftware
NOTE
The X-Series Signal Analyzer Calibration Files are located on the disk drive.
Therefore, replacement of the disk drive will invalidate the current instrument
calibration if the original calibration file has not been backed-up and restored.
However, a full calibration is required to assure that the instrument meets all
specifications. Arrangements regarding the level of calibration must be made
between the end user and the calibration provider.
Adjustments Required
All (if the Calibration file could not be backed-up).
None (if the Calibration file was backed-up and restored to the E:\ drive).
Performance Testing Recommended
All (if the Calibration file could not be backed-up).
None (if the Calibration file was backed-up and restored to the E:\ drive).
End of installation.
For assistance, contact your nearest Agilent Technologies Sales and Service Office. To find your local
Agilent office access the following URL, or if in the United States, call the following telephone
number:
http://www.agilent.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
Installation Note
N9000-90030