FPGA Code Update
There is FPGA program code on many different assemblies inside of the instrument, and all of these
must be a compatible version with the instrument software revision. In order to ensure that this
requirement is met you need to run an FGPA update program. If this program detects that there are
assemblies that need to be updated it will updated them to the correct version.
NOTE
Once you start the FPGA programming process you MUST NOT interrupt the
process for any reason. This would include turning the instrument off or
unplugging the power cord to the instrument. Doing so will result in an inoperative
instrument, requiring the affected assembly to be replaced.
Updating the Instrument FPGA Code 1.Connect a USB mouse and keyboard to the instrument
front panel USB ports.
C:\Program Files\Agilent\SignalAnalysis\Physics
2. Using the mouse, click Start, Log Off, Log Off.
3. Relogin as "administrator". The default password is "agilent4u".
4. Using the mouse, click on Start and select Explore to start Windows Explorer. Navigate to the
following folder:.
C:\Program Files\Agilent\SignalAnalysis\Physics
5. In this folder find and execute the file named:
FPGA_Prog.bat
6. The FPGA Programming Utility will start and a window as shown in
Figure 15
FPGA Programming Utility
7. To program the FPGA code enter 1 and press
pressing
and
1
Enter
8. The programming of the FPGA code could take a few minutes to complete. Once it has finished
the instrument will reboot itself to use the new code.
Installation Note
N9000-90030
Option PC3 - Single Core High Performance Processor Upgrade Kit
one more time.
. You will need to confirm this selection by
Enter
Figure 15
will appear.