Keysight CXA N9000A Installation Note page 16

Signal analyzer option pc3 single core high performance processor
Hide thumbs Also See for CXA N9000A:

Advertisement

Option PC3 - Single Core High Performance Processor Upgrade Kit
6. Signal Analyzer Application Pre-Run Configuration
In order for the signal analyzer application to startup and run properly there are two instrument
configuration tasks that need to be performed before you initially run the application. These steps
include:
License Key Installation
FPGA Code Update
In order to properly perform these tasks you will first need to stop the signal analyzer application from
loading. To do this follow the procedure below before proceeding to the next step in this installation
procedure.
1. When you see the Anti-Virus Message window shown in
external keyboard and select Task Manager from the Windows Security screen.
Figure 11
Anti-Virus Message Window
16
Figure 11
press CTRL+ALT+DEL on the
Installation Note
N9000-90030

Advertisement

loading