Download Print this page

Keysight CXA N9000A Installation Note page 6

Signal analyzer option pc3 single core high performance processor
Hide thumbs Also See for CXA N9000A:

Advertisement

Option PC3 - Single Core High Performance Processor Upgrade Kit
2. Calibration Data Backup
In order for the instrument being upgraded to not need all of the instrument adjustments and
performance verification tests to be run after this upgrade is completed the calibration data will need
to be backed up onto an external drive prior to removing the existing processor / disk drive assembly.
NOTE
While the backing up and restoring of the calibration data to the new disk drive
will return the instrument performance to the state that it was prior to this upgrade
this does not guarantee that all instrument performance parameters meet their
specified values. A full instrument calibration would be required to verify this.
Follow the steps outlined below for your calibration data backup.
1.
Connect a USB keyboard and mouse to two of the instrument rear panel USB ports.
2. Connect a USB storage device that has been scanned for virus to one of the front panel USB
port.
3. Press
System
4. The Alignment Data Wizard window as shown in
Figure 1
Alignment Data Wizard
5. Follow the on-screen instruction to backup the calibration data to the USB storage device.
The file will be saved with a .bak file suffix.
6
,
,
Alignments
Backup or Restore Align Data
.
Figure 1
will appear.
Installation Note
N9000-90030

Advertisement