Option PC3 - Single Core High Performance Processor Upgrade Kit
7. Calibration Data Restoring (SQL Database)
1. After the previous step of this upgrade procedure the instrument should have just rebooted and
once again have the Anti-Virus message window as shown in
2. Press
on either the instrument front panel or the external keyboard to continue, and the
Enter
signal analyzer application will now load.
3. Locate the USB storage device that was earlier used to backup the instrument calibration file and
insert it into one of the instrument front panel USB ports.
4. Once the signal analyzer application has completed loading restore the instrument calibration file
by pressing:
System
window as shown in
5. Follow the on-screen instruction to restore the previously saved calibration data from the USB
storage device.
No matter how the instrument calibration data was backed up prior to this upgrade, any further
backing up of the data will now be done through the instrument's Alignment Data Wizard.
6. Cycle the instrument power for the restored data to be used.
20
,
,
Alignments
Backup or Restore Align Data..
Figure 1
will appear.
Figure 11
displayed.
. The Alignment Data Wizard
Installation Note
N9000-90030