7 Position of Subject and Operator

1)
Position of subject (initial contact)
1a) Optimal position
1b) Tolerated position
1c) Not tolerated position / buffer zone
2)
Intended position of operator
The operator must be within reach of the emergency off at all times.
If the operator is not able to reach the emergency off button at the UserTerminal (body height, obstacles, etc.), the operator
must install an additional emergency stop within reach (see accessories).
3)
Training area
4)
Free area
5)
Safety area
6)
Patient environment
There must be no electrical devices within this area, which are not part of an ME-System with the device.
Do not touch the subject and external electrical devices at the same time.
© 1988 - 2019 h/p/cosmos sports & medical gmbh
40%, front
30%, middle
30%, rear
acc. to ISO 20957-1
acc. to ISO 20957-1
acc. to DIN EN 957-6
acc. to IEC 60601-1
running area
subject + device
training area + 0.6 m
2.0 m behind device
device + 1.5 m
cos105000_170-190_65, rev. 1.03
must be free at all times
(except operator)
must be free at all times
(except operator)