Alternating between inspiration and expiration, the
pressure gauge should indicate a consistent
inspiratory pressure and the preset end-expiratory
pressure of approximately 5 cmH
The display should now shows a minute ventilation of
CMV MV = 6.0
with a tolerance of ±1.0 L/min.
Disconnect test lung.
Display after approximately 25 seconds:
CMV MV = 0.0
Paw low
together with an intermittent audible alarm.
Reconnect test lung.
The intermittent audible alarm tone should stop after
approximately 25 seconds.
g
1 Press »
Reset« to clear the display.
If values are outside tolerance:
Replace flow sensor.
Checking End-Expiratory Pressure (PEEP)
2 Set »PEEP« knob to 0 cmH
3 Reading on pressure gauge at end of expiration:
0 cmH
O +2 cmH
O tolerance.
2
2
2 Set »PEEP« knob to 10 cmH
3 Reading on pressure gauge at end of expiration:
O ±2 cmH
10 cmH
O tolerance.
2
2
2 Set »PEEP« knob back to 0 cmH
O.
2
O.
2
O.
2
O.
2
Checking Readiness for Operation

Checking End-Expiratory Pressure PEEP

1
D
Oxylog 2000
D
Oxylog 2000
3
2
25