2. Refer to
Figure
14. Pull the AIF/DIF assembly part way out of the chassis. Disconnect cable
W15 from the analyzer.
3. Completely remove W15 from the analyzer.
Figure 14
AIF/DIF Assembly Partial Removal
4. Locate W36 in the Opt HL3 Cable Kit. This cable will have part number 8121-1862 and will be
labeled "902" on one end and "7" on the other. Connect the end labeled "902" to A15J902 on
the A15 EFEC Assembly. Connect the other end to A13J7 on the A13 Front End Assembly.
Torque the connector on A13J7 to 10 inch-pounds.
5. Locate W37 in the Opt HL3 Cable Kit. This cable will have part number 8121-1401 and will be
labeled "900" on one end and "100" on the other. Connect the end labeled "900" to A15J900
on the A15 EFEC Assembly. Route the cable along the left side of the instrument above the
fans. Route the cable down behind the rear of the second (rearmost) fan. Connect the cable to
A2J100 on the A2 EAIF Assembly
6. Slide the AIF/DIF assembly into the slot at the rear of the instrument and push on the
assembly to mate the connectors to the motherboard assembly.
Installation Note N9010-90066
Installation Procedure
23