Installation Procedure
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the signal analyzer
C:\Program Files\Agilent\licensing
Verify the License Installation
1. Before the licenses will be recognized, the XSA application must be restarted. Press System,
Exit Program. An Exit Analyzer dialog box will appear; press Enter to confirm the exit.
2. Double-click on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to
finish starting (the analyzer should be sweeping).
3. Press System, Show System on the analyzer to display a list of installed options. You should
see the following options listed ("x" is 1 for an EXA and 2 for an MXA).
— N90x0B-CR3 Connector Rear, 2nd IF Output
— N90x0B-CRP Connector Rear, Arbitrary IF Output