Installation Procedure
Update FPGA Code
On the first boot-up after installing the new hardware, the analyzer will detect the new A15 Front
End Controller. However, the FPGA code in the new A15 will not necessarily be synchronized with
the instrument software installed earlier. The following procedure ensures that the FPGA code in
the new A15 is updated to the correct revision.
1. Turn on the analyzer and wait for the XSA application to launch.
2. Close the instrument application software by pressing System, Exit Program, Enter.
3. Connect a USB mouse and keyboard to the instrument front panel USB ports.
4. Using the mouse, click Start, Shutdown, Log Off.
5. Relogin as "administrator". The default password is "Keysight4u!". Click OK.
6. Using the mouse, double-click on the Windows Explorer link on the desktop and navigate to
the following folder:
C:\Program Files\Keysight\SignalAnalysis\Physics
7. In this folder find and execute the file named:
FPGA_Prog.bat
8. The FPGA Programming Utility will start and a window as shown in
Figure 22
FPGA Programming Utility
9. To program the FPGA code enter 1 and press Enter. You will need to confirm this selection by
pressing 1 and Enter one more time.
10.The programming of the FPGA code could take several minutes to complete. Once it has
finished the instrument will reboot itself to use the new code and this process will then be
completed.
32
Figure 22
will appear.