Installation Procedure
2. Refer to
Figure
7. Pull the front frame carefully away from the chassis. Remove the ribbon
cable W1 from the A8 Motherboard.
Figure 7
Front Panel Cable
3. If the analyzer has Option BBA (BBIQ inputs), there will be an additional cable (W24) that will
need to be removed. W24 connects to the A18 BBIQ Interface Board. Pull the Front Frame
assembly carefully away from the chassis. Remove the ribbon cable W24 from the A18 BBIQ
Interface Board.
16
W1 may have locking springs on each side. Depress the spring on each side of the connector to
remove from the motherboard.
Installation Note N9010-90066