Standards Compliance; Rtu Communication Specifications; Cc-Link Communication Specifications; Input/Output Specifications - Mitsubishi Electric Super-S ME96SS Series Manual

Electronic multi-measuring instrument
Table of Contents

Advertisement

‡Standards Compliance

Electromagnetic Compatibility
Emissions
Radiated Emission
Conducted Emission
Harmonics Measurement
Flicker Meter Measurement
Immunity
Electrostatic discharge Immunity
Radio Frequency Electromagnetic field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
Conducted Disturbances, Induced By Radio Frequency Fields Immunity
Power Frequency Magnetic Field Immunity
Voltage Dips and Short Interruptions
Safety
Europe
U.S. and Canada
Installation Category
Measuring Category
Pollution Degree
®
‡MODBUS

RTU Communication Specifications

Item
Interface
Protocol
Transmission method
Connection type
Baud rate
Data bit
Stop bit
Parity
Address
Distance
Max. connectable units
Terminal Resistance
Recommended Cable
 For more information on data, please refer to the following document.
· Electronic Multi-Measuring Instrument ME series MODBUS

‡CC-Link Communication Specifications

Item
No. of occupied stations
CC-Link version
Baud rate
Transmission method
Synchronous method
Encoding method
Transmission path format
Transmission format
Error control system
Number of connectable units
Remote station numbers (station numbers)
 For CC-Link connection cables, please use the dedicated cables.
For information regarding dedicated cables, please refer to the CC-Link Partner Product Catalog published by the CC-Link Partner Association or CC-Link Partner Product
Information on the CC-Link Partner Association website (http://www.cc-link.org).
Notes 1. Dedicated CC-Link cables compatible with Ver. 1.00 cannot be used in tandem with dedicated CC-Link high-performance cables compatible with Ver. 1.00.
Notes 2. In the case of systems consisting of units compatible with Ver. 1.00, 1.10 or 2.00 used in tandem with Ver. 1.00 or 1.10 cables, Ver. 1.00 specifications will
apply for the maximum total cable length and length of cables between stations.
Notes 3. For terminal resistance, be sure to use 110 Ω ±5% (1/2W product) when using dedicated CC-Link cables or 130 Ω ±5% (1/2W product) when using dedicated
CC-Link high-performance cables.
 For more information on data, please refer to the following document.
· Electronic Multi-Measuring Instrument programming manual (CC-Link)...................................................LEN080334
· Electronic Multi-Measuring Instrument programming manual (CC-Link)(For ver. 2 remote device station)...LEN130391

‡Input/Output Specifications

Item
4-20mA (0~600 Ω)
Analog output
Pulse/Alarm output
No-voltage "a" contact Capacity: 35VDC, 0.1A
Digital input
19-30VDC 7mA or less
Digital output
No-voltage a contact Capacity: 35VDC, 0.2A
CE, as per EN61010-1
cRUus as per UL61010-1, IEC61010-1
III
III
2
RS-485 2-wire half-duplex transmission
RTU (binary data transfer)
Asynchronous
Multi-point bus
2400, 4800, 9600, 19200, 38400bps
8
1, 2
ODD, EVEN, NONE
1 to 255 (0: for broadcast mode)
1,200m (max)
31 units
120Ω 1/2W
Shielded twisted-pair AWG24 to 14
®
Interface specifications...LMS-0492
1 Station Remote device station
CC-Link Ver 1.10 / Ver 2.00
10Mbps / 5Mbps / 2.5Mbps / 625kbps / 156kbps
Broadcast polling system
Frame synchronous system
NRZI
Bus format (EIA RS485)
HDLC
CRC (X
+ X
+ X
+ 1)
16
12
5
42 units (max, remote device station)
1 to 64
Specification
EN61326-1/CISPR 11,
FCC Part15 Subpart B Class A
EN61326-1/CISPR 11,
FCC Part15 Subpart B Class A
EN61000-3-2
EN61000-3-3
EN61326-1/EN61000-4-2
EN61326-1/EN61000-4-3
EN61326-1/EN61000-4-4
EN61326-1/EN61000-4-5
EN61326-1/EN61000-4-6
EN61326-1/EN61000-4-8
EN61326-1/EN61000-4-11
Specification
Specification
Optional Plug-in Module type
ME-4210-SS96
ME-4210-SS96
ME-4210-SS96, ME-0040C-SS96, ME-0052-SS96
ME-0052-SS96
12

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents