Opening The Menu; Reading Out Information - Dräger Polytron 7000 Instructions For Use Manual

Type p3u and type p3fb
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Use
Key
OK
4.4.2
Passwords
Calibration password
Settings password
For setting the passwords, see section 9.3.
4.4.3

Opening the menu

● To open the Information directly: In measuring mode, press
for approx. 1 second.
● To open the Information and Calibration menus:
a. In measuring mode, press
b. Select Enter password.
c. Enter the calibration password.
● To open all menus:
a. In measuring mode, press
b. Select Enter password.
c. Enter the settings password.
4.4.4

Reading out information

In the menu Information choose the appropriate menu item:
Instrument
Function
Confirms inputs.
Selects menus and functions.
Access to
– Information menu
– Calibration menu
Factory setting: _ _ _ 1
Access to
– Information menu
– Calibration menu
– Settings menu
Factory setting: _ _ _ 2
Show Notice
Shows existing warnings. For troubleshooting with respect
to the warning number displayed
(e.g. #58) see section 7.2.
If multiple warnings are present, this is indicated (e.g. 1/3 =
page 1 of 3.).
Show Fault
Shows existing errors. For troubleshooting with respect to
the error number displayed (e.g. #100) see section 7.1.
If multiple errors are present, this is indicated (e.g. 1/3 =
page 1 of 3.).
Show Fault
Shows present error and warning codes in a table. If all
Codes
numerical groups are displayed with "00", no error is pres-
ent.
and hold down for approx. 3 seconds.
M
and hold down for approx. 3 seconds.
M
and hold down
M
Dräger Polytron 7000

Hide quick links:

Advertisement

Table of Contents
loading