Dräger Polytron 7000 Instructions For Use Manual page 32

Type p3u and type p3fb
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Troubleshooting
Number
# 51
# 52
# 53
# 58
# 59
# 64
# 106
# 111
# 112
# 114
# 115
# 119
# 120
# 131
# 132
# 135
Cause
Data logger in hold mode
(Stack) is filled to 100 %. It
is no longer recording
data.
Data logger in hold mode
(Stack) is filled to 90 %.
No valid date or time is set. Set time and date (see section 9.4).
Dongle was removed with-
out having been deacti-
vated.
Hardware fault of a dongle. Replace the dongle.
The pump is worn
Flow rate of the pump
module undershoots the
warning threshold. Reli-
able measurements will
soon no longer be possi-
ble.
Zero deviation too high.
Sensor not operated in the
specified temperature
range.
Sensor near end of life.
Calibration interval
expired.
The sensor was exposed
to an excessively high gas
concentration for a long
period.
The sensor is not yet fully
warmed up. Measuring
errors can occur.
The sensor was exposed
to an excessively high gas
concentration for a long
period.
The sensor will soon no
longer be ready for opera-
tion.
Sensor electrolyte has
evaporated. Measure-
ments will soon no longer
be possible.
Information, e.g. the part
number or the serial num-
ber, is not available.
Remedy
Read out the data, clear and restart
the data logger (see section 12).
Read out the data as soon as possi-
ble, then clear and restart the data
logger.
Deactivate Dongle prior to removing
(see section 9.9).
Replace the pump module.
Check hose for blockage, re-adjust
the flow rate, if necessary.
Perform zero adjustment (see
section 6.3).
Operate sensor in the specified tem-
perature range.
Replace the sensor (see section 8.2).
Calibrate the sensor again (see
section 6).
Reduce excessive gassing.
Wait until the sensor has fully warmed
up.
Reduce excessive gassing. If the
problem persists, replace the sensor
(see section 8.2).
Replace the sensor (see section 8.2).
Replenish sensor electrolyte.
Disconnect the gas detector from the
mains and restart it.
If the error persists, have the gas
detector checked by DrägerService.