Control Panel - Dräger Polytron 7000 Instructions For Use Manual

Type p3u and type p3fb
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Use
4.1.5
Checking the display for proper functioning
In order to recognise defective pixels in the display, it is possible to invoke the
Show Fault Codes menu image.
The menu image is suited since it shows many contiguous pixels, making it
possible to recognise line, column or individual pixel faults. Displayed fault codes
are not relevant in this context.
1. Open the menu (see chapter 4.4.3).
2. In the menu, select Information > Instrument > Show Fault Codes.
3. Check if all pixels are displayed.
4.2

Control panel

1
2
3
4.3
Info mode and function key
4.3.1
Info mode
In measuring mode, the Info mode can additionally be activated. It provides an
overview of device information presented on multiple pages.
Example display
Page 1
07.11.2011
SW Version : 8.0
Part No.
Serial No.
DeviceCode: 00006317
Instrument Info
Page 2
Sensorname: O
Part No.
Serial No. : XXXXXXX
EEPROM Typ : 1
EEPROM Vers.: 1
Sensor Info
Polytron
1
OK
2
M
Up key or function key
Down key or menu key
OK key
12:34
: 8317778
: ARUA0001
1/4
2
: 6809630
2/4
3
Description
Device information
Line 1 – Date and time
Line 2 – Software version
Line 3 – Device part no.
Line 4 – Device serial no.
Line 5 – Device code
Sensor information
Line 1 – Sensor name
Line 2 – Sensor part no.
Line 3 – Sensor serial no.
Line 4 – EEPROM type
Line 5 – EEPROM version
Dräger Polytron 7000

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents