Activating Info Mode; Using The Function Key; Menu Navigation - Dräger Polytron 7000 Instructions For Use Manual

Type p3u and type p3fb
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Example display
Page 3
Gasname
Range
4–20 SP
Alarm A1
Alarm A2
Sensor Config.
Page 4
Power
Fault
Warning :
Op.time
Pump info
1) Displayed only if a relay module is used!
2) Displayed only if a pump module is used!
4.3.2

Activating Info mode

● In measuring mode, press
4.3.3
Navigation in Info mode
Key
M
OK
After 30 seconds without key operation, the gas detector automatically exits the Info
mode.
4.3.4

Using the function key

The function key can be used to directly invoke a function that has been assigned to
the function key. By default, the error log is displayed.
● In measuring mode, press
For configuring of the function key (see section 9.7 Configuring the function key).
4.4
Menu
4.4.1

Menu navigation

Key
M
: O
2
: 25.00 Vo
: 25.00 Vo
: 19.00 Vo
: 23.00 Vo
3/4
2)
:
XXX %
:
X.X l/min
X.X l/min
:
XXXX h
4/4
Function
Scrolls to the next page.
Scrolls to the previous page.
Exits Info mode.
Function
Navigates upward.
Sets values.
Navigates downward.
Sets values.
Dräger Polytron 7000
Description
Sensor settings
Line 1 – Gas name
Line 2– Measuring range (cannot be edited) and
unit of measurement
Line 3 – Measuring range for the analogue interface
Displayed only if the analogue interface card is
installed
Line 4 – Alarm threshold A1 and unit
Line 5 – Alarm threshold A2 and unit
Pump information
Line 1 – Pump output
Line 2 – Flow threshold for error
Line 3 – Flow threshold for warning
Line 4 – Running time of the pump
and hold down for approx. 3 seconds.
and hold down for approx. 1 second.
Use
1)
1)