Setting user-definable parameters

Set temperature:
Fig. 15
Defrost time:
Fig. 16
24
(LED 1 and LED 3 on)
• (Press and hold „P" button and simulta-
neously press
Press the two buttons once again to deac-
tivate the defrost function.
During normal operation, the actual
cryochamber temperature is dis-
played.
• Press
once.
• Press
or
to select the desired tem-
perature.
• Press
once more.
• Press
or
value.
The first two digits correspond to the hours,
the last digit indicates minutes. Since there is
only one digit indicating minutes, the indicated
value must be multiplied by 10. Defrost time is
selectable in 10-minute steps.
Fig. 16 shows defrosting time set at 23:50
hours.