Fitting New Pump - Weinmann ACCUVAC Basic Service And Repair Instructions

Aspirator
Hide thumbs Also See for ACCUVAC Basic:
Table of Contents

Advertisement

4. Remove tube
from vacuum regulator.
58
5. Unscrew the 2 cylinder-head screws
remove the faulty vacuum regulator.
6. Fit the new regulator unit
7. When replacing the knob, ensure that the gap
between knob and keyboard membrane (with
the spindle screwed in) is approx. 1.0 mm.
8. Use a 10 mm tubular wrench to tighten the nut
securing knob
.
44
9. Insert cap
(slit in cap must match ridge in
45
knob).
10. Close the device (see "8.3 Closing the de-
vice" on page 13).
11. Perform a functional check (see "6.2 Perform-
ing the functional check" on page 8).

8.12 Fitting new pump

The pump is only available as a complete replacement unit.
Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see "8.2 Opening the de-
vice" on page 12).
2. Detach silicone fabric tube
pump.
3. Detach rolled-up silicone tube
connection.
4. Unscrew 4 screws
41
5. Detach tube
from faulty vacuum pump, at-
57
tach it to outlet of new pump
round motor.
6. Finish fitting the new vacuum pump in reverse
order.
7. Close the device (see "8.3 Closing the de-
vice" on page 13).
8. Perform a functional check (see "6.2 Perform-
ing the functional check" on page 8).
and
47
in reverse order.
46
from inlet port of
56
from muffler
57
from support.
and wind it
44
58
46
57
44
56
41
Repairs: Information and Instructions
47