Installation
1
Place a microcell in the Microcell Holder (if the cell has ground glass sides or
sides featuring the Cary logo, hold the cell by these sides) and check that the
cell aperture is vertically centered in the cell holder aperture.
2
If the cell aperture is not at the correct height, remove the cell and adjust the
adjustment screw in the Microcell Holder accordingly using the hexagonal ball
driver (2.5 mm). Replace the cell in the Microcell Holder.
3
Slide back the sample compartment lid. See Figure 4.
1
2
Figure 4. Cary 60 showing the sample compartment lid (1) and front door (2)
4
Remove the sample compartment front door. See Figure 4.
5
If not already fitted, install the cell holder base in the sample compartment as
follows:
a
b
6
Place the Microcell Holder on the cell holder base by aligning the holes in the
Microcell Holder over the raised black knobs in the cell holder base.
7
Tighten the thumbscrew on the Microcell Holder.
You should now align the Microcell Holder.

Alignment

To align the Microcell Holder:
1
Double-click Align in the Cary WinUV folder located on the computer desktop.
2
Select the Cary tab.
Agilent Cary 60 Spectrophotometer User's Guide
Place the cell holder base in the sample compartment, aligning the two
locating holes over the two locating pins in the floor of the sample
compartment.
Firmly tighten the thumbscrew on the cell holder base (item 2 in picture
above).