For optimum analytical performance, it is recommended that the ambient
NOTE
temperature of the laboratory be between 20 and 25 °C and be held constant to
within
The area should have a dust-free, low humidity atmosphere. Air conditioning is
recommended. The room should be temperature-controlled if your analyses are
particularly sensitive.
Condition
Non-operating (transport)
Operating within performance
specifications

Weights and Dimensions

Weight
Packed
Unpacked
Dimensions (width x depth x height)
Packed
Unpacked
The Cary 60 has been designed to withstand 10 kg, the approximate weight of a
17 in. LCD monitor.
The workbench should be about 90 cm (36 in) high. Remember to provide space
for the computer, monitor and printer.
To avoid damage through spillage of solutions and samples being analyzed, the
worktops should be covered with a material that is corrosion resistant and
impervious to liquids.
Allow at least two inches of space on both sides, and six inches at the rear of the
system to permit free air circulation. Power cord and all other connections are
located at the rear of the instrument. The Power switch is located on the front
panel.
16
2 °C throughout the entire working day.
±
Altitude (m, ft)
0–4600, 0–15000
0–3100, 0–10000
23 kg (51 lb)
18 kg (40 lb)
595 x 710 x 350 mm (24 x 28 x 14 in)
477 x 567 x 196 mm (19 x 23 x 8 in)
Temp. (°C, °F)
-40 to 70, -40 to158
5–40, 41–104
Agilent Cary 60 Spectrophotometer User's Guide
Specifications
Humidity (%RH)
non-condensing
<90
15–80