Cleaning The Test Cell - Ametek CHANDLER ENGINEERING 5265 Instruction Manual

Static gel strength analyzer
Table of Contents

Advertisement

pressure at temperatures above 212 ° F (100 ° C) may cause water in the test cell
to become steam.
2. Turn the PUMP WATER switch to the OFF position.
3. Turn MAIN POWER and INSTRUMENT POWER switches OFF.
4. Close the PRESSURE RELEASE valve (clockwise). Failure to do so will
result in water leakage when the U-tube or thermocouple is loosened.
5. Remove the U-shaped high-pressure tubing connecting the test cell to the
high-pressure filter on the autoclave assembly.
6. Disconnect the top sensor cable.
7. Disconnect the bottom sensor cable from the front panel.
8. Remove the thermocouple or disconnect the cable.
9. Lift the test cell from the autoclave assembly. Guide the bottom transducer
cable through the front panel of the instrument.

Cleaning the Test Cell

When the test cell has been cooled and removed from the autoclave, it should be
cleaned according the following guidelines.
1. Place the test cell in a vice, topside up. Use the wrench flats and do not
scratch or nick the cylinder or plugs.
2. Remove the top plug and pour off any standing water on the cement sample.
3. Remove the cell from the vice and replace in the vice topside down.
4. Unscrew and remove the bottom plug of the test cell.
5. Turn the cell over and drive the cement sample out of the test cell with a
hammer.
6. Clean the cement and grease from the top and bottom plugs and cylinder with
solvent.
7. When all traces of cement have been removed, grease the inner surfaces of the
test cell, including the seal and O-rings.
8. Replace the O-rings if they were damaged during the test.
9. The instrument is now ready to run another test.
SECTION 2 – OPERATING INSTRUCTIONS
2-9

Advertisement

Table of Contents
loading

Table of Contents