SECTION 2 – OPERATING INSTRUCTIONS
2-3
coolant switch in the OFF position to avoid possible temperature oscillations at
temperatures above room temperature.

Pressure Release Valve

Used to manually release pressure from the test cell. Valve must be closed during
testing except when it is necessary to manually release pressure. Valve must also
be closed when removing test cell with cooling water ON or else a significant
water leak will occur.

Cylinder Pressure Gauge

Displays the pressure inside the test cell.

Temperature Controller

Used to control the temperature in the test cell.

Heater Switch

Used to turn the flow of current to the heater ON or OFF. Switch must be in the
ON position during testing and should be in the OFF position as a safety
precaution at other times.

Main Power Switch

Used to turn the 220VAC power to the heater and internal solenoid valves
instrument ON or OFF. Switch must be in the ON position during testing.

Instrument Power Switch

Used to turn the power to the internal instrumentation ON or OFF. Switch must
be in the ON position during testing. The instrument power may be connected to
an uninterruptible power supply (UPS) if desired.

Left Rear Panel

The left rear panel contains all the connections for the
cables that connect the autoclave to the processor.
This panel also contains connections for the top
transducer, the thermocouple, the main power, and the
fuses.

Data Cable Connector

This connection is not used on the 5265 Autoclave.

Ref Cable Connector

This connection is not used on the 5265 Autoclave.