Unpacking And Installation - Leica HistoCore Arcadia C Instructions For Use Manual

Cold plate
Hide thumbs Also See for HistoCore Arcadia C:
Table of Contents

Advertisement

4.3

Unpacking and installation

4
HistoCore Arcadia C – Cold Plate
When the instrument is delivered, check the tilt indicators on the packaging.
If the arrowhead is blue, the shipment was transported laying flat, was tilted at
too great an angle or fell over during transport.
Note this on the shipping documents and check the shipment for possible damage.
1
2
3
Fig.3
4
Fig.4
4.
Setting up the instrument
These unpacking instructions only
apply if the box is placed with the
symbols
1. Remove the packing strap (1) and the adhesive
tape (2).
2. Open the package. Lift up and remove the
carton wall (3).
3. Remove the foam pads (4) one by one.
The HistoCore Arcadia C always has
to be transported upright and horizon-
tally. It must not be inverted under any
circumstances, even for short periods,
or stored on one of its sides.
It is mandatory to observe a wait-
ing time of 4 hours between the
last transport and the first time the
instrument is switched on. The oil
present in the compressor needs
this time to flow back to its original
location.
facing upwards.
15