2.
Tap the
[More]
button.
3.
Tap the [Length
Unit]
4.
Tap the desired length unit.
To activate or deactivate the connection tone:
The instrument is in Tier 1 Test mode.
1.
Press the [A] key.
2.
Tap the
[More]
button.
3.
Tap the [Enable
Tone]
connection tone.
To recall stored measurement data:
For detailed information, see "Memory Management" on
page 66.
To recall information relating to built-in sources:
The instrument is in Tier 1 Test mode.
1.
Press the [A] key.
2.
Tap the
[More]
button.
3.
Tap the
[Info]
button.
Changing settings for the Tier 1 loss test
Here, the setting can be made to add new, edit, or delete test
setups. Each Test Setup can be saved under its own name. The
factory setting is named Default.
Note: Test setups can only be created, deleted, or edited if the
instrument is configured as a Local instrument.
The basic setting in Tier 1 loss test is the Test Type. The Test
Type ca be set to Loopback or Remote. Depending on your
instrument hardware configuration the device ca be configured
for single-mode or multi-mode measurements. In Loopback
mode you need only one instrument configured as Local. The
measured cable pair needs to be looped with a jumper cable on
the far end while in Remote mode the far end needs to be
terminated with an OLTS-85(P) configured as Remote.
button.
button to activate or deactivate the
OLTS-85