Disturbance Analyzer (Click) Measurements
Making a Measurement
Step
7. If desired, set the
number of clicks as a
limit for data
collection
8. Set the Factor f of the
measurement
9. Collect the
Disturbance data
10. Review the final
analysis
Keysight N6141A Measurement Guide
Action
Click Count to
a. Press
select On.
b. Enter the desired value
using the numeric keypad
Enter
and select the
factor f and select
a. Press
one of the configured
Factor f menu keys.
Sweep/Control
a. Press
Start
.
Or,
Restart
b. Press
.
a. Automatically presented
after the data collection
has finished (either by test
duration or click count)
Notes
For more information see
page 41
.
key.
For more information see
page 42
.
The collected information is displayed automatically
,
into the disturbance list and categorized either as a
click or a continuous disturbance.
After the data collection has finished, the
Disturbance Measurement will automatically
analyze the data, apply all appropriate exceptions
(as defined in CISPR14) and present the results in
the final analysis window.
"Click Count" on
"Factor f" on
Collecting data
35