Keysight Technologies X Series Measurement Manual page 16

Signal analyzers, emi receiver
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Conducted Emissions Measurement Examples
Making Conducted Emission Measurements
Running Frequency Scan
Step
1. Turn on the DUT
2. Set Up
3. Start the scan
4. Stop the scan
16
Action
a. Configure the LISN and DUT.
b. Turn the DUT on.
Connect the DUT and LISN, to the EMI Receiver
as shown below:
Meas Setup
a. Press
,
Sequence
Scan Only
,
Sweep/Control
Start
,
Stop
a. Press
.
Notes
Ensure that the power cord
between the DUT and the LISN is
as short as possible. The power
cord can become an antenna if
allowed to be longer than
necessary.
Scan
Signals above the limit are
designated in red.
,
.
This step will not be necessary if
the measurement has completed
the number of scans set or the
desired time.
Keysight N6141A Measurement Guide