Keysight Technologies X Series Measurement Manual page 19

Signal analyzers, emi receiver
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Conducted Emissions Measurement Examples
Making Conducted Emission Measurements
Measuring the Quasi-peak and average values of the signals
Step
1. Perform a Re-measure on
all signals in the list
2. Review the measurement results
Measurement tip
If the signals you are looking at are in the lower frequency range of the
conducted band, 2 MHz or lower, you can reduce the stop frequency to get a
closer look. Note that there are fewer points to view. You can add more data
points using the scan table. The default setting in the scan table is two data
points per BW or 4.5 kHz per point in this case since the resolution bandwidth
is 9 kHz. To get more data points, change the points per bandwidth to four
points.
Keysight N6141A Measurement Guide
Action
Meas Setup
a. Press
,
Sequence
(Re)measure,
,
(Re)measure
All Signals
,
Sweep/Control
Start
,
Notes
Scan.
,
.
The delta to Limit Line values
should all be negative. If some of
the measurements are positive,
there is a problem with conducted
emissions from the DUT.
19