Keysight Technologies X Series Measurement Manual page 23

Signal analyzers, emi receiver
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Radiated Emissions Measurement Examples
Making Radiated Emission Measurements
Step
2. Turn on the
instrument.
3. Select the EMI mode.
4. Open the scan table
and select the desired
range
5. Set the attenuation
and internal amplifier
6. Load limit lines
7. Load correction
factors for the
biconical antenna
Keysight N6141A Measurement Guide
Action
a. Press the front-panel power
key.
Mode, EMI receive
a. Press
Meas Setup
a. Press
,
Table
.
Press Select Range
b.
Range 3, Range
to
More 1 of 3
a. Press
,
Attenuation
0
dB
,
,
,
Internal Preamp
Low
,
Band
.
Recall
Data
Limit
a. Press
,
,
(Preloaded 1)
Open
,
b. Scroll to EN 55022, Rad,
Class A, 30 MHz to 1
GHz (10m).csv, Open.
Amplitude
a. Press
Corrections (Preloaded 1),
Open
.
b. Select Antenna,
Biconical (30 MHz to
300 MHz).csv, Open.
Notes
This is the default mode.
r.
Scan
Deselect any range that has a green check.
,
On
.
.
23

Advertisement

Table of Contents
loading