Wiring the channels
February 2020
4.8.2
Wire the mA Input (externally powered)
Procedure
Wire to the appropriate input terminal and pins.
Figure 4-24: mA Input wiring (externally powered)
7
A
8
A. mA Input
B. 100 Ω input resistance at Channel D
C. 4–20 mA Input device
D. 30 VDC (maximum)
4.9

Wire the Discrete Input

Use this section to wire the Discrete Input in explosion-proof, nonincendive, or
nonhazardous installations.
Important
Meter installation and wiring should be performed only by suitably-trained personnel.
4.9.1
Wire the Discrete Input (internally powered)
Procedure
Wire to the appropriate input terminal and pins.
Figure 4-25: Discrete Input wiring (internally powered)
A
A. Discrete Input
B. Channel C or D
C. Switch
36
B
B
C