5.2.8.6 Maintenance Pressure
A pressure sensor on the 130143
PCB measures the pressure at the
UV block. When the pressure rises
above the threshold pressure, the
analyzer will automatically stop the
pumps, preventing the carrier gas
system from being damaged by the
corrosive liquids in the analyzer.
5.2.8.7 Fault Pressure

5.2.9 Language Screen

UVTOCSetup and System Configuration.fm
The IR gas channel has five filter settings to produce different
moving averages:
Setting
Sampling mode
Low
4 samples every 2 seconds
Medium
20 samples every 10 seconds
High
40 samples every 20 seconds
Very high
80 samples every 40 seconds
Adaptive
Adaptive filter
For this analyzer, Adaptive is the preferred setting.
1. To change the filter setting, press ENTER.
2. Asterisks will flash on either side of the displayed setting. When the
desired filter setting is displayed, press ENTER to store the selection in
memory.
This is the threshold above which the maintenance event for reactor pressure
will occur.
M ai nt Pr es s ur e
6 0 kA
This is the threshold above which the fault event for reactor pressure
will occur.
F au lt Pr es s ur e
10 0 kP a
The Language screen is entered by pressing the UP button while in the
Tolerances screen.
L an gu a ge
English
1. To change the displayed language, press ENTER.
2. Asterisks flash on either side of the displayed language.
3. Press the UP or DOWN buttons to scroll through the language selections.
The choices are English, Francais, Deutsch, Espanol, and Italiano.
Section 5
Page 85
Setup Menu