5. Once gas is applied, press and hold the R button to start the calibration. The 60 second
countdown will commence (refer to
monitor will be calibrated to the predefined values.
6. The timer will count up to 9 seconds, during which time the monitor will adjust the gain to the
sensors accordingly. No manual input is required.
7. Once calibration is successfully completed, the user has the option of saving the results (refer to
Figure 59: CAL COMPLETE/SAVE RESULTS
Figure 59: CAL COMPLETE/SAVE RESULTS Screen
8. Using the L button, highlight one of the 3 the available options :
A. UPDATE – saves the calibration data in the memory and automatically updates the
calibration due date based on the option stored in the monitor set-up menu.
B. SAVE ONLY – saves the calibration data in the memory but does not update the
calibration due date.
C. ABANDON – neither saves the calibration data in the memory nor updates the
calibration due date.
9. Press and hold the R button to accept this option