Configured Options Table
Option
Description
NEOD Warning
Sets the Near End Of Delivery warning time, as time left to End Of Delivery.
EOD Point
Sets the End Of Delivery point.
The Alaris Enteral Syringe Pump will switch to running at the KTO rate when EOD is
KTO Rate
reached.
VTBD Max
Maximum permissible Volume To Be Delivered value.
Display Syringe Brand
Displays the Syringe Brand on the main display.
Auto Save
Feature to retain previous settings when Alaris Enteral Syringe Pump is switched on.
AC Fail
The AC Power Failure Alarm will sound if the AC power is disconnected.
Bolus Mode
Bolus feature is set to be Hands-On or Hands-Free.
Pressure Display
Displays the Pressure Icon on the main display.
Bolus Rate Default
The default value for bolus rate.
Bolus Rate Max
The maximum value for bolus rate.
Bolus Volume Max
The maximum permissible bolus volume.
Cap Pressure
Sets the maximum pressure value.
Pressure Default
Sets the default occlusion alarm level.
Cap Rate
Sets the maximum value for delivery rate.
Purge Rate Max
Sets the purge rate.
Purge Volume Max
Sets the maximum permissible purge volume.
Call Back Time
The time for the Alaris Enteral Syringe Pump to sound the call back alarm.
Event Log Display
The event log can be displayed.
Battery Icon
Displays the Battery Icon on the main display.
Audio Volume
Sets the alarm volume of the Alaris Enteral Syringe Pump at high, medium or low.
Auto Night Mode
Backlight dims between hours 21:00 and 06:00.
1000DF00601 Issue 7
Alaris™ Enteral Syringe Pump
Default
10 minutes
1%
0.1ml/h
1000ml
10ml/h
200ml/h
1ml
L10
L4
200ml/h
500ml/h
2.0ml
2 minutes
Low
21/34
Configured Options