CareFusion Alaris Directions For Use Manual page 11

Enteral syringe pump
Hide thumbs Also See for Alaris:
Table of Contents

Advertisement

Do not mount the Alaris Enteral Syringe Pump with the AC power inlet pointing upwards. This could affect the
w
electrical safety in the event of a fluid spill.
Pole Clamp Installation
Never mount the pump using the pole clamp and remove the pump from service for examination by qualified
w
service personnel when:
• the pole clamp assembly shows signs of excessive wear
• the pole clamp pivot mechanism in the extended, mountable position has excessive loose movement
The pole clamp is fitted to the rear of the Alaris Enteral Syringe Pump
and will provide secure fixing to vertical drip poles of a diameter
between 15 and 40 mm.
1. Pull the folded pole clamp towards you and unscrew the clamp to
leave enough room for the size of the pole.
2. Place Alaris Enteral Syringe Pump around pole and tighten screw
until the clamp is secured to the pole.
Ensure the pole clamp is folded away and stored
w
within the recess at the rear of the Alaris Enteral
Syringe Pump before connecting to a Docking Station/
Workstation* or when not in use.
Never mount the Alaris Enteral Syringe Pump such that
the stand becomes top heavy or unstable.
Docking Station/Workstation* or Equipment Rail Installation
The rotating cam can be fitted to the rectangular bar on the Docking Station/Workstation* or the equipment rail measuring 10 by
25mm.
1. Align the rotating cam on the rear of the Alaris Enteral Syringe Pump with the rectangular bar on the Docking Station/Workstation*
or the equipment rail.
2. Hold the Alaris Enteral Syringe Pump horizontally, push the Alaris Enteral Syringe Pump firmly onto the rectangular bar or
equipment rail.
3. The Alaris Enteral Syringe Pump should click into position when fitted to the bar.
4. Ensure that the Alaris Enteral Syringe Pump is positioned securely.
5. To release, push the release lever and pull the Alaris Enteral Syringe Pump forwards.
* Alaris Gateway Workstation and Alaris DS Docking Station
Rectangular bar
Release lever (push to release)
1000DF00601 Issue 7
Alaris™ Enteral Syringe Pump
Recess
*
Rotating cam
10/34