Configured Options

This section comprises a list of configurable options, which can be entered via the Alaris Enteral Syringe Pump configuration menu
(available in Technician Mode).
Enter the access code on the Alaris Enteral Syringe Pump for Configured Options, see the Technical Service Manual for details.
Access codes should only be entered by qualified technical personnel.
w
Clock Set
1. Select CLOCK SET from the Configured Options menu using the
f
keys to adjust the date displayed, pressing the NEXT softkey to access the next field.
2. Use the
3. When the correct time and date are displayed press the OK softkey to return to the Configured Options menu.
Language
This option is used to set the language of messages shown on the Alaris Enteral Syringe Pump display.
1. Select LANGUAGE from the Configured Options menu using the
f
2. Use the
keys to select the language.
3. When the desired language has been selected press SELECT softkey to return to the Configured Options menu.
Contrast
This option is used to set the contrast on the Alaris Enteral Syringe Pump display.
1. Select CONTRAST from the Configured Options menu using the
f
2. Use the
keys to select a contrast ratio value. The contrast of the display will change when scrolling through the numbers.
3. When the desired value has been reached press the OK softkey to return to the Configured Options menu.
General Options
1. Select GENERAL OPTIONS from the Configured Options menu using the
2. Select the option required to enable/disable or adjust and press the MODIFY softkey.
3. When all the required modifications have been carried out press the QUIT softkey.
4. Either select the next configuration option from the menu or turn the Alaris Enteral Syringe Pump off, returning it to operation as
required.
NURSE CALL FITTED
NURSE CALL INVERT
RS232 SELECTED
QUIET MODE
Enables Nurse Call (hardware option).
When enabled, the nurse call output is inverted.
Sets the Alaris Enteral Syringe Pump's communications to use RS232 (hardware option).
The NURSE CALL FITTED option must be enabled to allow RS232 to be enabled.
Mode to silence key press tones and power down sequence.
1000DF00601 Issue 7
f
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys and press the OK softkey.