3 WARNINGS AND PRECAUTIONS

CAUTION
User's manual
NOTE
NOTE
CAUTION
NOTE
LASER RADIATION -
DO NOT STARE
INTO BEAM
1mW
CLASS II
635nm
LASER
PRODUCT
AVOID
EXPOSURE-
Laser radiation
is emitted from
this aperture.
LBL-X-100
NOTE
NOTE
NOTE
CAUTION
WARNINGS AND PRECAUTIONS
It is very important that the place where the
unit is to be used and the position from which
the user is to operate the unit are correctly
shielded. Since radiation safety requirements
vary from country to country and state to state
it is the responsibility of the user to ensure
that all local safety requirements are met.
If the x-ray unit has been stored at tempera-
tures under +10°C for more than a few hours,
time must be allowed for the unit to reach room
temperature before turning it on.
This X-ray unit may be dangerous to both
patient and operator unless safe exposure
values are used and correct operating proce-
dures are observed.
The patient positioning lights are class II laser
products (21 CFR § 1040.10).
Class 1 laser product (Standard EN 60825-
1:1994).
The patient positioning lights are class 1, in-
herently safe laser lights.
EMC requirements have to be considered, and
the equipment must be installed and put into
service according to the specific EMC informa-
tion provided in the accompanying docu-
ments.
Portable and mobile RF communications
equipment can affect Planmeca Proline EC X-
ray unit.
The remote exposure control, Autoprint and
Admark film marking systems are connected
to the X-ray units Planet connectors. Do NOT
connect any other devices to the Planet con-
nectors.