ERROR CODE
Er 62
Limit sensor power missing. Turn off the unit first for 30 seconds, try again without the
patient. If the error is still present, please contact your service technician.
Er 64
The radiation reaching the AEC sensor is very low or missing completely. Check that
the cassette is AEC compatible, i.e. marked with text "AEC COMPATIBLE". If the cas-
sette type is correct, contact your service technician.
Er 65
The radiation reaching the AEC sensor is too high. If the cassette is in the cassette
carriage and the patient is positioned when this error occurs, contact your service
technician.
Er 70
Processor communication error. Turn off the unit first for 30 seconds, try again without
the patient. If the error is still present, please contact your service technician.
Er 71
Generator processor program memory failure. Turn off the unit first for 30 seconds, try
again without the patient. If the error is still present, please contact your service techni-
cian.
Er 72
Keyboard processor program memory failure. Turn off the unit first for 30 seconds, try
again without the patient. If the error is still present, please contact your service techni-
cian.
Er 73
Keyboard processor program error. Turn off the unit first for 30 second. If the error is
still present, please contact your service technician.
Er 74
Keyboard processor operating improperly. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 75
Keyboard processor operating improperly. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 80
Keyboard processor EEPROM failure. Turn off the unit first for 30 seconds. If the error
is still present, please contact your service technician.
Er 81
Generator processor EEPROM failure. Turn off the unit first for 30 seconds. If the error
is still present, please contact your service technician.
Er 82
Keyboard processor configuration error. Turn off the unit first for 30 second. If the error
is still present, please contact your service technician.
Er 83
Generator processor configuration error. Turn off the unit first for 30 seconds. If the
error is still present, please contact your service technician.
Er 84
Tube power generator time out. Turn off the unit first for 30 seconds, try again without
the patient and the film cassette. If the error is still present, please contact your service
technician.
Er 90
Keyboard processor EEPROM not activated. Turn off the unit first for 30 seconds. If
the error is still present, please contact your service technician.
Er 91
Keyboard processor stack overflow. Turn off the unit first for 30 seconds. If the error is
still present, please contact your service technician.
Er 99
Error code generation error. Turn off the unit first for 30 seconds. If the error is still
present, please contact your service technician.
User's manual
ERROR MESSAGE EXPLANATION
Planmeca Proline EC Panoramic X-ray
TROUBLESHOOTING