Exposure warning indicator
Milliampere (mA) setting keys and display
Auto key
Control key
Clock key
User's manual
The exposure warning light will come on when you
press the exposure button to take an exposure. You
will also hear a warning signal. The tone of the warning
signal can be changed if required. See section "11.5
Adjusting the tone of the exposure warning signal" on
page 33.
During the exposure preparation time a low-pitched
signal is heard. A high-pitched signal indicates that the
unit is radiating.
The milliampere setting minus (-) key decreases the
10
milliampere value and the plus (+) key increases it.
mA
The maximum value is 12 mA and the minimum value
is 4 mA.
The auto key allows you to use certain programmed
parameter values. The parameter values are stored in
AUTO
the memory with auto and P keys. See section "11.7
Quick exposure settings" on page 34.
Some of the keys have second functions which can
only be used when the control key is pressed to acti-
CTL
vate the additional function mode. When this key is
pressed the indicator light will come on. This key is
also used to clear error codes from the display, see
section "14 TROUBLESHOOTING" on page 40.
Press the clock key once and the day and month will
replace the current time on the main display. Press the
key a second time and the year will appear. Press the
key a third time, or wait five seconds, and the current
time will reappear. If you wish to set the clock see sec-
tion "11.3 Setting the clock" on page 32.
CONTROL PANEL - KEYS AND DISPLAYS
Planmeca Proline EC Panoramic X-ray