4. Save the setting by pressing the
confirmation key. The month is
automatically highlighted. You can
change it with the turn control.
Accordingly, you can also set:
Year
Hour and minute
Time zone GMT (e.g. +1 in Germany, see Fig. 25 )
Daylight savings
-11 -10
-11 -10
-9
-9
-8
-8
-7
-7
Fig. 25
Time zones

7.5 Adjustment

The appliances are temperature calibrated and adjusted at the factory. In case readjustment
should be necessary later on – for example due to influence of the chamber load – the
appliance can be calibrated customer-specifically using three calibration temperatures of your
choice:
CAL1
Temperature calibration at low temperature
CAL2
Temperature calibration at medium temperature
CAL3
Temperature calibration at high temperature
For temperature adjustment, you will need a calibrated reference measuring device.
-6
-6
-5
-5
-4
-4
-3
-3
-2
-2
-1
-1
0
0
+1
+1
+2
+2
+3
+3
+4
+4
+5
+5
Menu functions
Date
27
.
05
Time
12 : 00
Time zone
GMT +01
Daylight savings
+6
+6
+7
+7
+8
+8
+9
+9
+10 +11 +12 -12-11
+10 +11 +12 -12-11
.
2012